产品概览
| 库存情况 | 需按需确认 |
| 分类 | Systems for the consumption measurement of compressed air & industrial gases |
| 保修 | 12个月(全新)· 3个月(二手/翻新) |
发货与库存
SD9000 何时可以发货?
SD9000 属于按需采购产品——请申请报价,我们会为这一具体单品确认准确的库存情况及发货时间。
可以亲自到店提取 SD9000 吗?
可以在我们位于荷兰格罗宁根的仓库安排自提——请在下单前联系我们,确认库存情况并预约提货时间。
是否提供快递服务?
可以——在为 SD9000 申请报价时告诉我们,我们会在回复中附上快递发货选项。
我们的工厂停产了,急需这个零部件——你们能帮忙吗?
如遇紧急故障,请直接致电 +31 (0)50-7110348 或发邮件至 sales@partnerplc.com 联系我们——我们可根据库存情况及目的地,为 SD9000 安排当日快递或专车加急配送。
Precise measurement of flow, consumption and medium temperature High accuracy, repeatability and measurement dynamics For detecting compressed air in industrial use With switching output, analogue output and pulse output Clearly visible 4-digit LED display Convenient operation via buttons or parameter setting via IO-Link
技术规格
Product characteristics
| Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 |
| Measuring range | 20...6835 l/min; 0.3...95.3 m/s; 1.5...410 m³/h |
| 20...6835 l/min | 0.3...95.3 m/s |
| Process connection | threaded connection R 1 1/2 external thread DN40 |
Application
| Application | for industrial applications |
| Media | compressed air |
| Note on media | air quality ISO 8573-1; class 141; class 344 |
| Medium temperature[°C] [°C] | 0...60 |
| Pressure rating | 16 bar; 1.6 MPa |
| 16 bar | 1.6 MPa |
Electrical data
| Operating voltage[V] [V] | 18...30 DC; (to SELV/PELV) |
| Current consumption[mA] [mA] | < 110 |
| Protection class | III |
| Reverse polarity protection | yes |
| Power-on delay time[s] [s] | 1 |
Inputs / outputs
| Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 |
Outputs
| Total number of outputs | 2 |
| Output signal | switching signal; analogue signal; pulse signal; IO-Link; (configurable) |
| Electrical design | PNP |
| Number of digital outputs | 2 |
| Output function | normally open / normally closed; (parameterisable) |
| Max. voltage drop switching output DC[V] [V] | 2 |
| Permanent current rating of switching output DC[mA] [mA] | 250; (per output) |
| Number of analogue outputs | 1 |
| Analogue current output[mA] [mA] | 4...20; (scalable) |
| Max. load[Ω] [Ω] | 500 |
| Pulse output | consumed quantity meter |
| Short-circuit protection | yes |
| Type of short-circuit protection | pulsed |
| Overload protection | yes |
Measuring/setting range
| Measuring range | 20...6835 l/min; 0.3...95.3 m/s; 1.5...410 m³/h |
| 20...6835 l/min | 0.3...95.3 m/s |
| Display range | 0...8200 l/min; 0...114.4 m/s; 0...492 m³/h |
| 0...8200 l/min | 0...114.4 m/s |
| Set point SP | 55...6835 l/min; 0.8...95.3 m/s; 3.5...410 m³/h |
| 55...6835 l/min | 0.8...95.3 m/s |
| Reset point rP | 20...6800 l/min; 0.3...94.8 m/s; 1.5...408 m³/h |
| 20...6800 l/min | 0.3...94.8 m/s |
| Analogue start point ASP | 0...5125 l/min; 0...71.5 m/s; 0...307.5 m³/h |
| 0...5125 l/min | 0...71.5 m/s |
| Analogue end point AEP | 1710...6835 l/min; 23.8...95.3 m/s; 102.5...410 m³/h |
| 1710...6835 l/min | 23.8...95.3 m/s |
| In steps of | 0.001 m³ |
| 5 l/min | 0.1 m/s |
| Pulse value | 0.005...4 000 000 m³ |
| Pulse length[s] [s] | 0,018...2 |
| Measuring range[°C] [°C] | 0...60 |
| Display range[°C] [°C] | -12...72 |
Accuracy / deviations
| Repeatability[% of the measured value] [% of the measured value] | ± 1,5 |
| Accuracy (in the measuring range) | ± (3 % MW + 0,3 % MEW) / ± (6 % MW + 0,6 % MEW); (class 141 /; class 344; conditions: installation to DIN ISO 2533; installation in pipes: DN40) |
| Accuracy[K] [K] | ± 2; (medium flow in the limit area of the flow measurement range) |
Response times
| Response time[s] [s] | 0.1; (dAP = 0) |
| Damping process value dAP in steps[s] [s] | 0 - 0,2 - 0,4 - 0,6 - 0,8 - 1 |
Software / programming
| Parameter setting options | Flow monitoring; quantity meter; Preset counter; Temperature monitoring; hysteresis / window; normally open / normally closed; current/pulse output; display can be rotated and switched off; Display unit; totaliser |
Interfaces
| Communication interface | IO-Link |
| Transmission type | COM2 (38,4 kBaud) |
| IO-Link revision | 1.1 |
| SDCI standard | IEC 61131-9 CDV |
| SIO mode | yes |
| Required master port type | A |
| Process data analogue | 3 |
| Process data binary | 2 |
| Min. process cycle time[ms] [ms] | 4.1 |
| Supported DeviceIDs | default; 269 |
| default | 269 |
Operating conditions
| Ambient temperature[°C] [°C] | 0...60 |
| Storage temperature[°C] [°C] | -20...85 |
| Max. relative air humidity[%] [%] | 90 |
| Protection | IP 65 |
Tests / approvals
| EMC | DIN EN 61000-6-2; DIN EN 61000-6-3 |
| CPA approval | model number; 002TG; accuracy class; -; maximum allowable error; ± 4 % FS; Q (min); 1,4 m³/h; Q (t); -; Q (max); 410 m³/h |
| model number | 002TG |
| accuracy class | - |
| maximum allowable error | ± 4 % FS |
| Q (min) | 1,4 m³/h |
| Q (t) | - |
| Q (max) | 410 m³/h |
| Vibration resistance | DIN EN 68000-2-6; 5 g (55...2000 Hz) |
| DIN EN 68000-2-6 | 5 g (55...2000 Hz) |
| MTTF[years] [years] | 227 |
| Pressure Equipment Directive | Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request |
Mechanical data
| Weight[g] [g] | 4004 |
| Housing | rectangular |
| Dimensions[mm] [mm] | 475 x 90 x 120 |
| Materials | PBT-GF20; NBR; PC; stainless steel (304/1.4301); PTFE; brass coated; aluminium powder-coated |
| Materials (wetted parts) | stainless steel (304/1.4301); FKM; ceramics glass passivated; PEEK GF30; polyester; aluminium |
| Process connection | threaded connection R 1 1/2 external thread DN40 |
Displays / operating elements
| Display | Display unit; 5 x LED, green (Nl/min, Nm³/h, Nm/s, Nm³, °C); function display; 1 x LED, green; switching status; 2 x LED, yellow; measured values; alphanumeric display, 4-digit; programming; alphanumeric display, 4-digit |
| Display unit | 5 x LED, green (Nl/min, Nm³/h, Nm/s, Nm³, °C) |
| function display | 1 x LED, green |
| switching status | 2 x LED, yellow |
| measured values | alphanumeric display, 4-digit |
| programming | alphanumeric display, 4-digit |
Remarks
| Remarks | MW = measured value; MEW = Final value of the measuring range; Measuring, display and setting ranges refer to the standard volume flow according to DIN ISO 2533.; For information about installation and operation please see the operating instructions. |
| Pack quantity | 1 pcs. |
Electrical connection
| Connector | 1 x M12; coding: A |
12个月保修适用于所有全新库存
3个月保修适用于二手及翻新零部件
30天退货未使用零部件退货免收上架费
全球发货全球快速可追踪配送